|
If you ship liquid diagnostic specimens by air, are you sure
that you are complying with the regulations? Hospitals, labs,
clinics, etc. ship thousands of specimens every day and all manner
of primary receptacles are used. There are some packages currently
on the market that the manufacturers claim are compliant with
IATA Packing Instruction 650. This regulation
dictates, in part, that either the primary or secondary container
used in a shipment must pass a 95 kPa pressure differential test.
Unfortunately, many of these shipping systems offer a secondary
container that does not meet these requirements. As a
consequence, your primary container must pass to
make the shipment legal.
Take a look at your shipping system. If in doubt, ask the
manufacturer if their vials pass the required test. If all your
primaries have been tested and certified, fear not. If they are
not designed to withstand this test, and the secondary container
you use does not qualify, your shipments do not comply with
the regulations.
With large operations, it is virtually impossible to standardize
all primary containers to be legal under PI 650. A far easier,
cost effective, and legal approach would be to seek out a packaging
system with a secondary container that meets or exceeds the prescribed
pressure test criteria.
A leaking package could result in a very high fine or civil
liabilities with no upper limit.
Regulations are written with the main purpose of safety. Abiding
by them should be your "primary" concern.
A better solution is to use a shipping system that uses a
pressure tested secondary container such as the STP-210
or the STP-250.
|